> Towards the end of run 5276 D5 tripped because of high > temperture. The magnet watch was able to reset it but > could not find the time of the trip because it is not > recorded well in the on screen error log of the magnet > control panel in building 1002B. > Since there is no chance of beam for 2 days I turned down > the magnets, set them to standby and went to bed. Two comments -- for shifters - the instructions for operating magnets (as in the magnet proecedure you have all read and signed) when leaving IR magnets should be turned first to standby then to off, and the key put back in the lockbox. -- To check for actual change in mag field one can either a) daq web page magnet status -- Database log of 'setpoints' b) Keithely readout of Hall probes that can make a dump i.e. listing of time + values recorded.