Date: Sat Sep 8 08:07:14 2001, Shift: 00:00-08:00
Supervisor: sjs, with: sjs
Log book#8 pages 94 - 95
Shift Summary:
------------------------------------------------------------
Long store beginning at 1:45 continues.
Run 4639. (1:55 - 2:35) Min. bias data run (trig 1,4, 8,
no downscaling). Stopped after ZM came in to turn on T3
anode voltages.
Run 4640. (2:35 - 6:55) Long min. bias run (trig 1,4,8,
no downscaling). 421,470 trig 1 events. Stopped to
start new run.
Run 4641. (6:58 - ...) Continuation of run 4640.
In scanning through SuperMon plots, noticed the following:
ZDC: -Times from CF discriminator--Only left time 1 and 3
present.
-High-gain ADC pedestals - No right pedestal 1
BB: -LeftBigADC - Right ADC9 gain low
H1: - H1PedDist - slots 13, 33 missing
-H1TOPDist - slot 15 missing
-H1BOTDist - slot 33 missing
TOFW: -TOFWPed - slot 56 low
-TOFWTDC - bot51-55 missing
TOFWCal: -TOFWCalADC top3 very broad pedestal
T4: -T4-1-Y2 - wire 110 low
T5: -T5-1-X - wire 7 missing
-T5-1-UPlanes - wire 11 missing
-T5-3-VPlanes - V1 wire 6 low; V2 wire 6 missing
Checking MaxADC per pad
TPM1: Row 8 pad 8-14 low
Row 10 pad 8-14 low
Row 11 pad 64-71 low (also low on reference plot)
Row 12 pad 64-71 low (also low on reference plot)
TPM2: Row 5 pad 40-47 low
Row 6 pad 40-47 low
ZM noticed a number of apparent card and/or cabling problems
with T3 that will require IR access.
------------------------------------------------------------
This archive was generated by hypermail 2b30 : Sat Sep 08 2001 - 08:07:43 EDT