We have calibrated runs with 1 dead ZDC module

From: Michael Murray (murray@CyclotronMail.tamu.edu)
Date: Thu Jun 20 2002 - 10:08:24 EDT

  • Next message: Claus O. E. Jorgensen: "reduction status"

    ----- Forwarded message from Andrei Makeev <makeev@comp.tamu.edu> -----
    Date: Thu, 20 Jun 2002 02:27:52 -0500
    From: Andrei Makeev <makeev@comp.tamu.edu>
    Reply-To: Andrei Makeev <makeev@comp.tamu.edu>
    Subject: one more e-mail to send...
    To: murray@cyclotronmail.tamu.edu
    
    Michael,
    
    could you please send my last e-mail to Flemming,
    which I also forwarded to you (if you didn't get it the
    contents is below).
    
    thanks,		Andrei.
    
    ==========================
    
    Hi Flemming,
    
    I commited the updated BrZdcRdoModule.cxx file to the CVS.
    
    For the runs where there is no signal in module1 of right ZDC (low-gain),
    I put (by hand) all slew paramteres for this module to be zero in the 
    calibration ASCII file. So, later, in BrZdcRdoModule I check for this
    condition (all slew pars for mod#1 are zeros), and if such, I calculate
    time from right ZDC using just 2-nd and 3-d modules.
    That was my first idea how to bypass that problem.
    
    Please let me know is Ok or not to do like this.
    
    Thank you,              Andrei.
    
    PS: Module seemed to work properly for calibrations for runs 5361-5368,
    which lack signal in right mod#1.
    
    ==========================
    
    ----- End forwarded message -----
    
    
    
    Michael Murray, Cyclotron TAMU, 979 845 1411 x 273, Fax 1899
    