¥To
study this effect we create huge signals by replacing the radiator with a
scintillator.
¥The
signals is split and used to START and STOP a TAC. The STOP is attenuated to ensure that the amplitude is below
the clamping amplitude.
¥The
time jump is observed during PR01 shows up when we use the 4413 CAMAC
discriminator.
¥No
time jump is observed when use the 623B NIM discriminator.