¥What do we learn from
this analysis?
¥If we
require a multiplicity of 1 in either array, we rule out contributions from
cross talk.
¥The
vertex resolution for all combinations of detectors is between 2.5 and 3.5
cm.
¥The
different detectors are clearly not aligned in time (a 1 ns shift is required
to move the vertex by 15 cm).
¥The
centroids of the vertex difference distributions can be used to realign the
individual T0 counters on either side.