3/18/2003 Crosstalk study

3/18/2003 Crosstalk study. 1

Notes about time calibrator runs. 1

Summary. 2

 

Last updated: 3/18/2003 1:51 PM

The plan:

Take 6 measurements:

11578 – all channels enabled

11579 – only T0P0 enabled in T0P, all other masked out in the LEDI discriminator.

11580 – all channels enabled

11581 – only T0P4 enabled in T0P,

11582 – all channels enabled

11583 – only T0P9 enabled in T0P,

11584 – all channels enabled

 

Run logbook comments:

Run Log Date Run Type Mag Events (seq) Totals Comments Details

11584  03/17 14:26:50  FBTIME  OFF  20743(1) 

 T0 special, all enabled.   

11583  03/17 14:23:43  FBTIME  OFF  20676(1) 

 T0 special. Enabled only T0P9 out of all TOP. Rates T0P=T0N=640, &=0   

11582  03/17 14:20:42  FBTIME  OFF  19647(1) 

 T0Special, all enabled   

11581  03/17 14:17:49  FBTIME  OFF  20137(1) 

 T0 Special. Enabled only T0P4 in T0P, rates T0p=T0N=680, T0N&T0P=0   

11580  03/17 14:14:52  FBTIME  OFF  19266(1) 

 T0 Special. All enabled.   

11579  03/17 14:12:17  FBTIME  OFF  19545(1) 

 T0 special. Only T0P0 enable in T0P. Rates T0N=T0P=660, T0N&T0P=0 Strange.   

11578  03/17 14:08:13  FBTIME  OFF  29629(1) 

 T0 special. All channels enabled. Note T0N counts 670 T0P = 1334. TON&TOP = 670   

 

 

Notes about time calibrator runs

The T0P&TON starts counting when at least 2 channels in T0P are enabled. (I checked this again 3/18 at 12:59 and T0P&TON now fires from single T0P)

The TOP is counting 2 times faster than T0N, this is true only when all 10 channels are enabled.

 

 

 

11578

All

11579

Only T0P0

 

 

11580

All

11581

Only T0P4

 

 

11583

Only T0P9

 

 

11584

All

Summary

1. It is confirmed that there is a significant positive time shift in (all) channels when signal is alone relative to the case when all signals on are fired.

T0P0 – pair on the edge. Shift = +38*25 ps

T0P4 – pair in the middle. Shift = +43*25 ps

T0P9 – last active pair out of 17 in the cable. Shift = +30*25 ps

2. The dependence of the shift from the pair position on the cable is not obvious.

3. The time drift between measurements was small ~ 2*25 ps.