| High Voltage To TOF | Look at HV_Control.vi on the Plastic computer. | ||
| Nominal TOF Status | Actual TOF Status | Agree? | |
| On c Off c | On c Off c | Yes c No c |
If they disagree, inform TOF Expert. |
| Instructions | Measurements | Comments | Actions To Take |
| TOF Discriminators | |||
| Go to the Camac controls VI. | Is X and Q valid? | Yes c No c |
If no, read out crates and recheck. If still invalid, note in logbook and restart vi. |
| Diagnostic Run Mode Checks | |||
| Go to the 4-fold logic unit in the TOF NIM bin. Look at the UPPER unit only. | Are channels A,B,C and D Off? (Upper part.) | Yes c No c |
If no, insert pin into pinholes A, B, C, and D if one isn't present. |
| The LED HV is located above the HV mainframe. | Is LED HV Off? | Yes c No c |
If No, dail down, counter clock wise, the HV to zero and disable HV. |
| Look at the Time Calibrator in the TOF NIM Bin. | Is Time Calibrator Off? | Yes c No c |
If No, switch off. |