TOF Expert Daily Check List
- Take Time Calibration Run, and run
FBGainCalc.
- Take FB Pedestal Run, and rn
FBPedCalc.
- Check that the ADC gate input on the Fastbus CAT module is
properly terminated.
- Check that the ICA Stop input on the Fastbus CAT module is
properly terminated.
- During normal running, check to see
that the Time Calibrator is on.
- Check that channel D is enable on the Upper
Section of the Logic Unit in the TOF NIM Bin in the electronics room.
- Check to see that the Delay Gate
Generator that enables the Time Calibrator is set to 10s Width on the Top
section and 10ms width on the bottom.
- Check that counter to make sure the
reference pulse generated by the time calibrator is counting around 3 Hz, or
generates ~3 counts for every enable pulse.
- Check the LED voltages and make
sure they are off and set to 0V for normal physics running.
- Run over a raw data sequence to
make sure the TOF is working properly. Run TOFMonitor_Setup.C then
TOFMonitor_SM.C to create a root file filed with numerous histograms. Then run
DisplayTOFMonitorHistos.C to view the histograms in the root file.