Gerrit van Nieuwenhuizen and Rachid Nouicer
Procedure made by : August 17, 2001
Shift Crew responsibility
- make sure that beams are stored and cogged
for 15 minutes without major
losses before ramping up silicon bias
- check radiation levels on chipmunks, should be less than 200 Microrad/Second.
If not don't switch on silicon, call Expert
Notes :
1 - Data taking for Stored and Cogged Beam (no
Steering)
: say the
following magic words :
- we are happy about
the collision rate (if the collision rate is reasonable)
- ask MCR what their
plans for this store are, will they steer the beam ?
- if they want to do
steering then don't ramp up the bias, wait until they have
finished steering and call them again
- tell them that you
would like to switch on silicon and ask them if they expect the
conditions to be
stable
- ask MCR to alert
PHOBOS 10 minutes before they dump the beam, so that we
can switch off the
silicon
- thank them for their
efforts
if FEC Low Voltage Power Supplies is already ON jump to Step 2.
if FEC LOW Voltage Power Supplies is OFF follow instruction :
if FEC Low Voltage Power Supplies is already on, use document "Magnet, General
Information/Help" for ramping the magnet.
- The Silicon FEC Low Voltage Power Supplies should be already ON,
DAQ is running without writing and no Latchup.
- For Latchup see latchup section in this document.
- Check the radiation levels on the Chipmunks again, should be less
than 200 Microrad/Second
If not, don't switch on silicon, switch off FEC Low Voltage, call expert..
- Ramp up the Silicon Bias by clicking on the green
button under the Bias Power Supplies Control on SiMon
- Click on <Restart ROCDB> under the <SiExpert> menu on the DAQ gui.
- Wait until the Silicon Bias has ramped up and all indicators are
green
- During the bias ramps, the Central Alarmer will go off, this is
normal
- After the Silicon bias turned on, one FEC (FEC 27) will have
latched up, and both the Latchup
window on the <OnLatchup1> and the <OnlLatchup 2> client will turn red, and
the <Central Alarmer > will sound
- Follow the instructions for
below
- Stop the run on the DAQGui
- Enable writing to Disk
- Wait for 5 minutes after the Si bias has been ramped up to give
the system time to stabilize
- Start a new run, reset scalers
- you are now taking data
2 - End of the Store
1) If dumping (or
quenching) of store by RHIC leads to the Latchup with the number
of ports with Vfp = 0 is less than or equal 8
:
1- Stop the run.
Start a new one without writing.
2- Click on the button
3- click on <Reset Latchup
under the <SiExpert> menu on the DAQ gui.
<OnLtchup1 (after 40 seconds) > and <OnLatchup2 (takes upto 2
minutes)
windows should be green
4- Click on <Restart ROCDB> under the <SiExpert> menu on the DAQ
gui.
5-Click on the Si hybrids on
the portal and check Vfp (Vfp should not be
equal to 0) and if Vfp not equal to 0 latchup Reset was successfull. You are ready for next
Store.
6- If latchup
persist Call Silicon Expert.
2) If dumping (or quenching )
of store by RHIC leads to the Latchup with
(Vfp =0) larger than 8
ports latched up :
1- Stop the run
2- Click on the button
and call Silicon Expert first.
- If the 8 ports or more
with Vfp = 0 appear in the FECs 17,18,19,20,21 and 22,
- click on
the button
on the bottom part of the screen
on SiMon .
- follow
instruction
- If you have general latchup
more than 8 ports with Vfp=0 including some
FECs from 17 to 22 :
2- Click button
on the bottom part of the screen on SiMon
3- Wait 10 minutes before switching
ON FEC Low Voltage Power Supplies
4 - In meantime you can ramp down
Phobos magnet
5 - follow instruction
- Stop the run
- Click on the button
- Don't TURN OFF the Low Voltage Power Supplies.
in case of power dip or storm (bad weather) or ...
- Stop the run
- Click button
on the bottom part of the screen on SiMon
3 - When Latchup happens, what should you do ?
- Initial state after ramping up the Silicon Bias
- Click on the Si hybrids on the portal which should not be
(old).
- After ramping up the bias the following
FEC ports will have latched up:
FEC Port
27
0&1&2&3
If more ports are latched up call a
Silicon EXPERT
- Click on <Reset Latchup> under the <SiExpert> menu on the DAQ
gui.
- Click on <Restart ROCDB> under the <SiExpert> menu on the DAQ gui.
- After 40 seconds, the <OnLatchup1> and
<OnLatchup2 ( need 2 minutes)>
windows should be green . Click on the Si hybrids on the portal and check
Vfp (Vfp should not be equal to
0), otherwise call Silicon EXPERT
- If you have a latchup, as indicated by the <OnlLatchup1> or <OnLatchup2>
display, you should:
- Stop the run. Start a new one without
writing.
- Click on <Restart ROCDB> under the <SiExpert>
menu on the DAQ gui.
- Click on the Si hybrids on
the portal which should not be (old).
- Record in the online logbook which, and how many,
FEC/Ports have
Vfp = 0 (last column) :
1 - If the number of ports with Vfp
= 0 is less than or equal to 8,
click on <Reset Latchup under the <SiExpert> menu on the DAQ gui.
<OnLtchup1 (after 40 seconds) > and <OnLatchup2 (takes
upto 2 minutes) windows should be green
2- If the number of
ports with Vfp = 0 is larger than 8, call
Silicon Expert
- If
the 8 ports or more with Vfp = 0 appear in the FECs 17,18,19,20,21 and
22,
- click
on the button
on the bottom part of the screen
on SiMon
.
- follow
instruction
- If you have general latchup
more than 8 ports with Vfp=0 including some
FECs from 17 to 22 :
2- Click button
on the bottom part of the screen on SiMon
3- Wait 10 minutes before switching
ON FEC Low Voltage Power Supplies
4 - In meantime you can ramp down
Phobos magnet
5 - follow instruction
- Initial state after switching on the silicon FEC Low Voltage Power
Supplies
- Click on <Restart ROCDB> under the <SiExpert> menu on the DAQ gui.
- Click on the Si hybrids on the portal which should not be
(old)
- After switching on the silicon FEC Low Voltage Power Supplies,
the following FEC ports will have latched
up:
FEC Port
9 2&3
17 3
18 3
21 2&3
27 2&3
If more ports are latched up call a
Silicon EXPERT
- Click on <Reset Latchup> under the <Expert> menu on the DAQ
gui.
- Click on <Restart ROCDB> under the <SiExpert> menu on the DAQ
gui.
- Click on the Si hybrids on the portal and check Vfp
(Vfp should be not equal to 0)
- After 20 seconds, the <OnLatchup1> and
<OnLatchup2 (need 2 minutes)> windows should be green, otherwise call Silicon
EXPERT
- Initial state after switching on the silicon FEC Low Voltage Power
Supplies
- Click on <Restart ROCDB> under the <SiExpert> menu on the DAQ
gui.
- Click on the Si hybrids on the portal which should not be
(old)
- After switching on the silicon FEC Low Voltage Power Supplies,
the following FEC ports will have latched
up:
FEC Port
17 3
18 3
21 2&3
If more ports are latched up call a
Silicon EXPERT
- Click on <Reset Latchup> under the <Expert> menu on the DAQ
gui.
- Click on <Restart ROCDB> under the <SiExpert> menu on the DAQ
gui.
- Click on the Si hybrids on the portal and check Vfp
(Vfp should be not equal to 0)
- After 20 seconds, the <OnLatchup1> and
<OnLatchup2 (need 2 minutes)> windows should be green, otherwise call Silicon
EXPERT
if FEC Low Voltage Power Supplies is already ON jump to Step 3.
if there is no steering and the beams are stable :
- Kill DAQGui, if it is running, by doing the following:
1) stop the run, if it is active
2) click <Quit PHATDAQ> under <File> menu
3) type <control-C> in the RUN
xterm
- switch on the silicon FEC Low Voltage Power Supplies by doing the
following (make sure lights in the tunnel are off) :
1) click green <ON> button in FEC PS Control on
SiMon
2) all the status buttons should appear as <ok>
and be green, with the
exception of 2 red
buttons under <OctVtx> (this is normal and will go
away in a few steps) .
The alarmer will sound, this is normal.
- start up the DAQGui by clicking on the PhatDAQRunControl icon on
the desktop. See also instructions for starting the DAQ.
- Wait until the DAQGui has appeared and the message "Ready to
receive initialization parameters"
has appeared in the <SiROC Messages> then you must click on the <Init
Silicon> item under the <Expert> menu
- Wait for the "waiting for CMD" message in the <SiROC
Messages> window
- Disable Writing to Disk
- Set runtype to AllBeam
- Start a run
- Click on the
icon on the menu bar at the bottom of the screen
on Online Monitoring 3, this will restart all online
clients
- Follow the instructions for :
above
- Follow the instructions for :
above
- The 2 red buttons on SiMon under <OctVtx> should turn green
within 20 seconds
- Click on <Restart ROCDB> under the <SiExpert> menu on the DAQ gui.
- The <OnlLatchup1> client on the Online Monitoring #3
computer should also become green in 30 seconds
- The <OnlLatchup2> client on the Online Monitoring #3
computer should also become green after few minutes
- If either the <OnLatchup1> or the <OnlLatchup2> client
becomes red, or the <OnlLatchup1> client alarm goes off on the <Central
Alarmer>,
(make sure they are running), call Silicon EXPERT.