Gerrit van Nieuwenhuizen and Rachid Nouicer
Procedure made by : August 17, 2001

    

    Shift Crew responsibility

Notes :


1 - Data taking for Stored and Cogged Beam (no Steering)

  :  say the following magic words :
            - we are happy about the collision rate (if the collision rate is reasonable)
            - ask MCR what their plans for this store are, will they steer the beam ?
            - if they want to do steering then don't ramp up the bias, wait until they have
               finished steering and call them again
            - tell them that you would like to switch on silicon and ask them 
               if they expect the conditions to be stable
            - ask MCR to alert PHOBOS 10 minutes before they dump the beam, so that we 
               can switch off the silicon
            - thank them for their efforts 
if FEC Low Voltage Power Supplies is already ON jump to Step 2. 
if FEC LOW Voltage Power Supplies is OFF follow instruction : 


if FEC Low Voltage Power Supplies is already on, use document "Magnet, General Information/Help"  for ramping the magnet.


2 -  End of the Store

1)  If dumping (or quenching) of store by RHIC  leads to the Latchup with the number 
      of ports with Vfp = 0 is less than or equal 8
           1- Stop the run. Start a new one without writing.
           2- Click on the button   
           3- click on <Reset Latchup under the <SiExpert> menu on the DAQ gui. 
                <OnLtchup1 (after 40 seconds) >  and <OnLatchup2 (takes upto 2 minutes) 
                windows should be green
           4- Click on <Restart ROCDB> under the <SiExpert> menu on the DAQ gui.
           5-Click on the Si hybrids on the portal and check Vfp (Vfp should not be 
               equal to 0) and if Vfp not equal to 0 latchup Reset was successfull. You are ready for next Store.
           6- If latchup persist Call Silicon Expert.

2)  If dumping (or quenching ) of  store by RHIC leads to the Latchup with 
       (Vfp =0) larger than 8 ports latched up :       
        1- Stop the run 
        2- Click on the button  and call Silicon expert
      - If the 8 ports or more with Vfp = 0 appear exclusively in the FECs 17,18,19,20,21 and 22, 
            - click on the  button on the bottom part of the screen 
               on SiMon . 
            - follow instruction
     - If you have general latchup more than 8 ports with Vfp=0 including some 
         FECs from 17 to 22  and call Silicon expert before proceeding to next step : 

        2- Click button on the bottom part of the screen on SiMon
        3- Wait 10 minutes before switching ON FEC Low Voltage Power Supplies
        4 - In meantime you can ramp down Phobos magnet
        5 - follow instruction


  in case of power dip or storm (bad weather) or ...


3 -  When Latchup happens, what should you do ?