Rachid Nouicer and Gerrit van Nieuwenhuizen
Procedure made by : January 20, 2003
Shift Crew responsibility
- make sure that beams are stored and cogged
for 15 minutes without major
losses before ramping up silicon bias
- check radiation levels on chipmunks, should be less than 1000
Microrad/Second.
If not don't switch on silicon
Notes :Important ! If, for some reason, more
than 10 FEC ports are latched up, do NOT issue a latchup
reset since this can damaged the power supplies of
silicon. [called Si expert].
Section 1 : Data taking for Stored and
Cogged Beam (no Steering)
: say the
following magic words :
- we are happy about
the collision rate (if the collision rate is reasonable)
- ask MCR what their
plans for this store are, will they steer the beam ?
- if they want to do
steering then don't ramp up the bias, wait until they have
finished steering and call them again
- tell them that you
would like to switch on silicon and ask them
if
they expect the conditions to be stable
- ask MCR to alert
PHOBOS 10 minutes before they dump the beam, so that we
can switch off the silicon
- thank them for their
efforts
if FEC Low Voltage Power Supplies is already ON jump to Step 2.
if FEC LOW Voltage Power Supplies is OFF follow instruction :
in the end of this documentation and after that you jump to Step 2 .
- The Silicon FEC Low Voltage Power Supplies should be already ON,
DAQ is running without writing and no Latchup.
- For Latchup see Section 3 (when Latchup happens) in this
document.
- Check the radiation levels on the Chipmunks again, should be less
than 500 Microrad/Second
If not, don't switch on silicon, switch off FEC Low Voltage, call expert..
- Ramp up the Silicon Bias by clicking on the green
button under the Bias Power Supplies Control on SiMon
- Click on <Restart ROCDB> under the <SiExpert> menu on the DAQ gui.
- Wait until the Silicon Bias has ramped up and all indicators are
green
- During the bias ramps, the Central Alarmer will go off, this is
normal
- After the Silicon bias turned on, one FEC (FEC 27)
will have latched up, and both the Latchup
(we are not sure, at this point if this latchup still occurs)
window on the <OnLatchup1> and the <OnlLatchup 2> client will
turn red, and the <Central Alarmer > will sound
- Follow the instructions for
below
- Keep DAQ running without writing for 5 minutes after the Si bias
has been ramped up to give the system time to stabilize
- Stop the run on the DAQGui
- Enable writing to Disk
- Start a new run,
- you are now taking data
Section 2 :
End of the Store
1) Stop the run
2) Wait until MCR has ramped down
our magnet
3) Switch
off Silicon with: Click button
on the bottom part of the screen on SiMon
- Stop the run
- Click on the button
in case of power dip or storm (bad weather) or ...
- Stop the run
- Click button
on the bottom part of the screen on SiMon
Section 3
: When Latchup happens, what should you do ?
- Initial state after ramping up the Silicon Bias
- Click on the Si Latchup Vfp Summary on the portal which
should not be (old).
- After ramping up the bias the following
FEC ports will have latched up:
FEC Port
27
0&1&2&3
If more ports are latched up
make a note in the logbook about which FEC/port are latched up.
(see
the important note in the beginning of the document)
- Click on <Reset Latchup> under the <SiExpert> menu on the DAQ
gui.
- Click on <Restart ROCDB> under the <SiExpert> menu on the DAQ gui.
- After 40 seconds, the <OnLatchup1> and
<OnLatchup2 ( need 2 minutes)>
windows should be green . Click on the Si Latchup Summary Vfp on the portal and check
that all indicator are green, otherwise call
Silicon EXPERT
- If you have a latchup, as indicated by the <OnlLatchup1> or <OnLatchup2>
display, you should:
- Stop the run. Start a new one without writing.
- Click on <Restart ROCDB> under the <SiExpert>
menu on the DAQ gui.
- Click on the Si Latchup Vfp Summary
on the portal which should not be (old).
- Record in the online logbook which, and how many,
FEC/Ports have
red indicator :
1 - If the number of ports with red indicator is less than
or equal to 10,
click on <Reset Latchup under the <SiExpert> menu on the DAQ gui.
<OnLtchup1 (after 40 seconds) > and <OnLatchup2 (takes
upto 2 minutes) windows should be green
2- If the number of ports latchup (red indicator on the
Silicon Latchup Vfp Summary) is larger than 10, call Silicon
Expert
- If the 10 ports or more
latchup (red indicator) appear in the FECs 1,17,18,19,20,21 and 22,
- click on the button
on the bottom part of the screen
on
SiMon .
- follow instruction
- If you have general
latchup more than 10 ports (red indicator) including some
FECs from 17 to 22 and
call Silicon expert before proceeding to next step :
1- Click button
on the bottom part of the
screen on SiMon
2- Wait 10 minutes before switching ON FEC Low
Voltage Power Supplies
3 - follow instruction
- Initial state after switching on the silicon FEC Low Voltage Power
Supplies
- Click on <Restart ROCDB> under the <SiExpert> menu on the DAQ gui.
- Click on the Si Latchup Vfp Summary on the portal which
should not be (old)
- After switching on the silicon FEC Low Voltage Power Supplies,
the following FEC ports will have
probably latched
up:
FEC Port
1
2&3
And possibly Also:
9 2&3
FEC Port
17 3
11 0
18 3
27 0
21 2&3
27 2&3
If more ports are latched up
make a note in the logbook about which FEC/port have latch up
(see the important
note in the beginning of the document)
- Click on <Reset Latchup> under the <Expert> menu on the DAQ
gui.
- Click on <Restart ROCDB> under the <SiExpert> menu on the DAQ
gui.
- Click on the Si Latchup Vfp Summary on the portal and
check Vfp (all indicators should be green )
- After 30 seconds, the <OnLatchup1> and
<OnLatchup2 (need 2 minutes)> windows should be green, otherwise call Silicon
EXPERT
- Initial state after switching on the silicon FEC Low Voltage Power
Supplies
- Click on <Restart ROCDB> under the <SiExpert> menu on the DAQ gui.
- Click on the Si Latchup Vfp Summary on the portal which
should not be (old)
- After switching on the silicon FEC Low Voltage Power Supplies,
the following FEC ports will have latched
up:
FEC Port
17 3
18 3
21 2&3
If more ports are latched up call a
Silicon EXPERT
- Click on <Reset Latchup> under the <Expert> menu on the DAQ
gui.
- Click on <Restart ROCDB> under the <SiExpert> menu on the DAQ
gui.
- Click on the Si Latchup Vfp Summary on the portal and
check Vfp (Vfp should be not equal to 0)
- After 20 seconds, the <OnLatchup1> and
<OnLatchup2 (need 2 minutes)> windows should be green, otherwise call Silicon
EXPERT
- Kill DAQGui, if it is running, by doing the following:
1) stop the run, if it is active
2) click <Quit PhatDAQ> under <File> menu
- DO NOT PERFORM THIS STEP IF THIS IS JUST FOR
switch on the silicon FEC Low Voltage Power Supplies by doing the
following (make sure lights in the tunnel are off) :
1) click green <ON> button in FEC PS Control on
SiMon
2) all the status buttons should appear as <ok>
and be green, with the
exception of 2 red
buttons under <OctVtx> (this is normal and will go
away in a few steps) .
The alarmer will sound, this is normal
- start up the DAQGui by clicking on the icon PhatDAQRunControl on
the desktop. See also instructions for starting the DAQ (click the
"OK" button).
- Wait until the DAQGui has appeared.
- Wait for the "waiting for CMD" message in the <SiROC
Messages> window
- Disable Writing to Disk
- Set runtype to AllBeam
- Start a run
- Click on the
icon on the menu bar at the bottom of the screen
on Online Monitoring 3, this will restart all online clients
- Follow the instructions for :
OR Follow the instructions for :
- The 2 red buttons on SiMon under <OctVtx> should turn green
within 20 seconds after resetting the latchup
- Click on <Restart ROCDB> under the <SiExpert> menu on the DAQ gui.
- The <OnlLatchup1> client on the Online Monitoring #3
computer should also become green in 30 seconds
- The <OnlLatchup2> client on the Online Monitoring #3
computer should also become green after few minutes
- If either the <OnLatchup1> or the <OnlLatchup2> client
becomes red, or the <OnlLatchup1> client alarm goes off on the <Central
Alarmer>,
(make sure they are running), call Silicon EXPERT.