Dear Bob and Flemming,

Apache wrote:

>3) T5-1 X1,X2,U1, U2 planes have dead wires.
>
I see the same. X1,X2 (wires 1-7) and U1,U2(wires 6-11) all go to the 
same mutliplexer
slot and are merged in to one MUX-TDC cable tagged with D10. My first 
guess it that
this is mux. board problem (broken fuse) or bad  mux-tdc connection.

U1,U2 (wires 1-5) is a selected problem. These channels go to the
same Ecl-Mux cable tagged with C3. In this case it might be a broken 
fuse on the
LV splitting board (if red diode is on).

I see that you successfully fixed the T4. Was it a broken fuse on mux board?

Best regards
Pawel.

>   The beam condition in E-log (when new run is 
>   started) is from jan 7.
>   The D4 WFG loopback seems to be stuck - the 
>   measured value is good.
>------------------------------------------------------------
>