Beam-Beam Commissioning

Using Summary of the data for the August Run, we picked best runs to analyze data for the Left and Right Arrays. Run 000217 for the Blue Ring (beam in the direction of the RA) and Run000231 for the Yellow Ring (beam in the direction of the LA). Relation between ADC Channel Number and Tube Number can be found here. TDC channels were numbered identically.

BLUE Beam:

Here are we have the most interesting pictures of Run 217:

YELLOW Beam:

Based on this analysis we can argue, that our ADC spectra can also be used as a rough probe for the centering of the beam.

Looking at the ADC spectra for different triggers one can see, that events which was triggered by Beam-Beam and Tiles trigger together have around 20% less intensity, than in case of a pure Beam-Beam trigger. Analysis of a trigger log book and real pulse pictures suggests that BB and Tiles trigger came earlier by about 25-40 nsec for the gate width of 70nsec. So, we made so simulations to see if this could be responsible for the 20% intensity drop. Here we have a pictures of some simplest simulations of this situation.

Counting Rates: Another interesting question would be, how do we have to set our trigger, so we don't start counting from the Beam-Gas collisions (accidental coincidences). In other words, coincidence of how many hits in both arrays do we have to consider to be sufficient for our trigger? Here you can find some simplest calculations and analysis of GEANT simulations, regarding this question.

Some calibrations were made for the Adc spectra:

   - only events with Beam-Beam Trigger were selected;
   - pedestal was subtracted;
   - Adc values were divided by the Gain value from the Lab studies for 
     every particular tube.

Another calibration were made with the Gain values from the Hamamatsu catalog. Only Lab Gain calibration gave us some reasonable results, such as, same energy values of the peaks for the neighboring tubes. Here are some results of the calibration with the Lab measured Gain values:

BLUE Beam:

YELLOW Beam:

All of the above proves, that our method of calibration is acceptable. Using it we hope to have some rough estimations of the Multiplicities on very early stages of analysis.


to the Beam-Beam Counters Page
Prof. Burton Budick, Yury Blyakhman.
Last modified: Wed Apr 25 17:46:17 EDT 2001